User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Ametek 1964:Digistrobe DATE: 30-May-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 16 NUMBER OF LINES: 139 CONFIGURATION: HP 5335A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 IEEE [@5335]IN 1.003 IEEE [@5335][TERM LF] 1.004 IEEE [@5335]FN1,AF1,MS60,GA1; 1.004 IEEE [@5335]TR1,AT.2; 1.005 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.006 DISP This procedure requires this use of the 1.006 DISP photovoltaic cell and the HP 5335A counter. 1.007 DISP Tape the photocell to the front of the UUT. 1.008 DISP Connect the output of the photocell to the 1.008 DISP INPUT A on the HP 5335A counter 1.009 HEAD {} 1.010 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍ»} 1.011 HEAD {º RPM Tests º} 1.012 HEAD {ÈÍÍÍÍÍÍÍÍÍÍͼ} 1.013 HEAD ÍÍ RPM Tests ÍÍ 1.014 JMP 2.001 1.015 EVAL 2.001 DISP Set the UUT to approximately 200 RPM. 2.002 MEMI Use keyboard to enter actual UUT reading in RPM. 2.003 MEME 2.004 IEEE [@5335][I] 2.005 MEME 2.006 MEMC ÿRPM 0.10% +1U #! Test Tol 0.2, Sys Tol 0.002, TUR 100.000 (>= 4.00). 3.001 DISP Set the UUT to approximately 500 RPM. 3.002 MEMI Use keyboard to enter actual UUT reading in RPM. 3.003 MEME 3.004 IEEE [@5335][I] 3.005 MEME 3.006 MEMC ÿRPM 0.10% +1U #! Test Tol 0.5, Sys Tol 0.005, TUR 100.000 (>= 4.00). 4.001 DISP Set the UUT to approximately 1000 RPM. 4.002 MEMI Use keyboard to enter actual UUT reading in RPM. 4.003 MEME 4.004 IEEE [@5335][I] 4.005 MEME 4.006 MEMC ÿRPM 0.10% +1U #! Test Tol 1, Sys Tol 0.01, TUR 100.000 (>= 4.00). 5.001 DISP Set the UUT to approximately 2000 RPM. 5.002 MEMI Use keyboard to enter actual UUT reading in RPM. 5.003 MEME 5.004 IEEE [@5335][I] 5.005 MEME 5.006 MEMC ÿRPM 0.10% +1U #! Test Tol 2, Sys Tol 0.02, TUR 100.000 (>= 4.00). 6.001 DISP Set the UUT to approximately 3000 RPM. 6.002 MEMI Use keyboard to enter actual UUT reading in RPM. 6.003 MEME 6.004 IEEE [@5335][I] 6.005 MEME 6.006 MEMC ÿRPM 0.10% +1U #! Test Tol 3, Sys Tol 0.03, TUR 100.000 (>= 4.00). 7.001 DISP Set the UUT to approximately 4000 RPM. 7.002 MEMI Use keyboard to enter actual UUT reading in RPM. 7.003 MEME 7.004 IEEE [@5335][I] 7.005 MEME 7.006 MEMC ÿRPM 0.10% +1U #! Test Tol 4, Sys Tol 0.04, TUR 100.000 (>= 4.00). 8.001 DISP Set the UUT to approximately 5000 RPM. 8.002 MEMI Use keyboard to enter actual UUT reading in RPM. 8.003 MEME 8.004 IEEE [@5335][I] 8.005 MEME 8.006 MEMC ÿRPM 0.10% +1U #! Test Tol 5, Sys Tol 0.05, TUR 100.000 (>= 4.00). 9.001 DISP Set the UUT to approximately 6000 RPM. 9.002 MEMI Use keyboard to enter actual UUT reading in RPM. 9.003 MEME 9.004 IEEE [@5335][I] 9.005 MEME 9.006 MEMC ÿRPM 0.10% +1U #! Test Tol 6, Sys Tol 0.06, TUR 100.000 (>= 4.00). 10.001 DISP Set the UUT to approximately 7000 RPM. 10.002 MEMI Use keyboard to enter actual UUT reading in RPM. 10.003 MEME 10.004 IEEE [@5335][I] 10.005 MEME 10.006 MEMC ÿRPM 0.10% +1U #! Test Tol 7, Sys Tol 0.07, TUR 100.000 (>= 4.00). 11.001 DISP Set the UUT to approximately 8000 RPM. 11.002 MEMI Use keyboard to enter actual UUT reading in RPM. 11.003 MEME 11.004 IEEE [@5335][I] 11.005 MEME 11.006 MEMC ÿRPM 0.10% +1U #! Test Tol 8, Sys Tol 0.08, TUR 100.000 (>= 4.00). 12.001 DISP Set the UUT to approximately 9000 RPM. 12.002 MEMI Use keyboard to enter actual UUT reading in RPM. 12.003 MEME 12.004 IEEE [@5335][I] 12.005 MEME 12.006 MEMC ÿRPM 0.10% +1U #! Test Tol 9, Sys Tol 0.09, TUR 100.000 (>= 4.00). 13.001 DISP Set the UUT to approximately 10000 RPM. 13.002 MEMI Use keyboard to enter actual UUT reading in RPM. 13.003 MEME 13.004 IEEE [@5335][I] 13.005 MEME 13.006 MEMC ÿRPM 0.10% +1U #! Test Tol 10, Sys Tol 0.1, TUR 100.000 (>= 4.00). 14.001 DISP Set the UUT to approximately 11000 RPM. 14.002 MEMI Use keyboard to enter actual UUT reading in RPM. 14.003 MEME 14.004 IEEE [@5335][I] 14.005 MEME 14.006 MEMC ÿRPM 0.10% +1U #! Test Tol 11, Sys Tol 0.11, TUR 100.000 (>= 4.00). 15.001 DISP Set the UUT to approximately 12000 RPM. 15.002 MEMI Use keyboard to enter actual UUT reading in RPM. 15.003 MEME 15.004 IEEE [@5335][I] 15.005 MEME 15.006 MEMC ÿRPM 0.10% +1U #! Test Tol 12, Sys Tol 0.12, TUR 100.000 (>= 4.00). 16.001 HEAD {} 16.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 16.002 DISP {º THIS COMPLETES THE VERIFICATION º} 16.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 16.003 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.